Description
The Pycnometer from Quantachrome is
designed to measure the true volume of solid materials by employing Archimedes’
principle of fluid displacement and gas expansion (Boyle's Law). Ideally, a gas
is used as the displacing fluid since it penetrates the finest pores assuring
maximum accuracy.
Technical Details
User Protocol
-
Turn the machine on and let it warm up for 1 hour
-
Adjust the nitrogen flow to 131kP (1.3bar)
-
Insert usb stick and chose for calibration and measurements that
data is written on usb.
-
Micro-Ultrapyc’s
Adapter Sleeves, Cells (micro, meso, nano) and Cover.
-
When pycnometer has not been used for 6
months, calibrate (press 2) Vadded:
-
Chose
to calibrate Vadded small
-
Insert
the micro cell (2nd from left on the picture)
-
Wait 10
min
-
Insert
the volume (1.0725)
-
Start
(first part of calibration takes 30min)
-
Insert
large calibration sphere
-
Wait 10
-
Start
(second part of calibration takes 30min)
-
The added
volume is around 1.66 cc
-
Calibrate every day Vcell (press 2):
-
Insert micro cell with one
large calibration sphere
-
Chose cell size small (enter)
--> Enter the volume (1.0725) (enter) --> multi-run (2) (enter) -->
max runs: 5 (enter) --> deviation:0.001 (enter) --> deviation: 0.001
(enter) --> print no (2) --> email no (2) --> usb yes (1) (enter)
-
Calibrate Vcell small (takes
10min)
-
Vcell is around 6.9 cc
-
To start a measurement press 1:
-
Weigh sample in micro cell
(micro cell should be filled at least ¾. For first samples check that
pycnometer works ok: measure density of sample when micro cell is filled for ¾
and when micro cell is filled completely).
-
Insert sample and close lid
-
measure (1) --> sample
parameter (2) --> weight: .... (enter) --> id: ... (enter) --> start
(3)