Protocol for safe use of the furnace
Principal
How to use the furnace
There are two possible ways to use the
furnace:
1. Only temperature controlled.
The furnace goes to the setpoint temperature and stays at this temperature until
it is stopped manually.
|
Press the page button to scroll through the menu
until SPºC appears.
|
|
Use the Up and Down keys to alter the value (ºC)
|
After setting the temperature the value
will be stored directly. The furnace will directly start heating.
To Stop: Switch of the power using the main
switch or use a lower setpoint temperature.
2. Timer Programmed.
Timer type t1
the timer waits for the setpoint temperature to be reached, then begins the
countdown. On completion of the countdown, the furnace switches the power to
the elements off (and flashes END on the display).
Setting up timer program
a.
|
|
Set the controller to Hold mode!
Hold the Up and Down keys together for 1.5 sec,
until the display flash HOLd
|
b.
|
|
Set the timer type to t1
Press and hold the Page key for 1.5 sec to enter the
menu
Press the Page key until ttYP shows on the display.
|
|
|
Use the Up and Down keys to set the timer type to t1
|
c.
|
|
Set the temperature setpoint
Press the Page key until SPºC shows on the display.
|
|
|
Use the Up and Down keys to alter the desired
temperature.
|
d.
|
|
Set the setpoint ramp rate
Press the Page key until SPrr shows on the display.
|
|
|
Use the Up and Down keys either to switch off or to
alter the rate value (ºC/min). Max is 30ºC/ min.
|
|
|
|
e.
|
|
Set the setpoint time
Press the Page key until t1
shows on the display.
|
|
|
Use the Up and Down keys either to switch off or to
alter the time value (hr:min).
|
f.
|
|
Start the timer
Press the Page key until SPrr shows on the display.
|
Some extra functions while the timer is
running:
|
|
Checking the remaining time
Press the timer key once.
|
|
|
Resetting the timer (only when timer count has ended
or the timer is paused)
Press and hold the Timer key for 1.5sec. Displays: rST to
indicate timer reset
|